Nellcor Puritan Bennett 840 PB840 Ventilator Medtronic Mechanical Ventilator PB 840 – Faulty

Sold!!!

Out of stock

SKU: 20032711-3714RM Category:

Description

Accessories – Air & O2 Hoses Only
This ventilator was purchased from a local government agency.
This ventilator is NOT in working order.
Upon testing the ventilator fails self test due to the below errors:
* FS0001 – O2 Flow Sensor Cross Check Failed (Q1)
* FS0003 – Air Flow Sensor Cross Check Failed (Q2)
* EST Passed.
* SST Failed: ‘Flow Sensor Calibration Failed’.
This could be a simple fix by replacing the Q1 & Q2 sensors [from USD $67.50 each.
Ventilator hours : 29,967.
Options: BiLevel, PAV+